Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.
Reason for recall: Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy. Distribution: Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines. Quantity: 12 US, 3 OUS Lot/code info: LifeShield Infusion Safety Software Suite v2.1 and v2.2 UDI-DI 00887709122802
00887709122802