When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.
Reason for recall: When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen. Distribution: Distribution was made to CA, FL, GA, IA, IL, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, OK, RI, SC, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Australia, Austria, Belgium, China, France, Germany, Italy, Japan, Jordan, Lebanon, Poland, Portugal, Romania, Russian Federation, Singapore, Spain, Switzerland, Taiwan, Thailand, and United Quantity: 169 systems Lot/code info: Model 870231: Software version 16.0.4, UDI (01)00884838099081(10)16.0.4.60330; Software version 16.4.2, UDI (01)0084838103054(10)16.4.2.60330, UDI (01)00884838102965(10)16.4.2.60330, UDI (01)00884838102972(10)16.4.2.60315, UDI (01)00884838102989(10)16.4.2.60330; UDI (01)00884838103191(10)16.4.2.60330; and Software version 16.4.3, UDI (01)00884838102965(10)16.4.3.60360. Model 870237: Software version 16.4.6, UDI (01)00884838103566(10)16.4.6.21197; and Software version 18.0.1, UDI (01)00884838103566(10)18.0.1.2149 and (01)00884838103566(10)18.0.1.21197.
00884838099081, 0084838103054, 00884838102965, 00884838102972, 00884838102989, 00884838103191, 00884838103566
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.