All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Reason for recall: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. Distribution: Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Ph Quantity: 95,704 Lot/code info: Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 770G Insulin Pump: 763000456634, 763000414344, 763000413750, 763000545604, 763000586188, 763000642242, 763000926113, 763000482602, 763000416553, 763000439903, 763000416546, 763000439910, 763000629892, 763000629908, 00763000595906, 763000412685, 763000418151, 763000416546, 763000439903, 763000545499, 763000595906, 763000545505, 763000629892, 763000642235, 763000414344, 763000413750, 763000439866, 763000413750, 763000586195, 763000545604, 763000578640, 76300058618802, 76300058619501, 76300057864002, 76300054560402, 76300058619502, 763000926113: For MiniMed 770G System Users Ages 2-13(M010831C001 1); MiniMed 770G SYSTEM USER GUIDE(M984970A002 1, M978700A002 1, M030281C001 3, M030281C002 1, M811470A261 2, M984970A004 1, M020090C001 3, M020090C002 1, M020090C003 1, M998590A012 2, M998590A003 1, M998590A004 1); MiniMed 770G SYSTEM USER GUIDE mmol/L(M000039C021 1, M811470A021 2).
763000456634, 763000414344, 763000413750, 763000545604, 763000586188, 763000642242, 763000926113, 763000482602, 763000416553, 763000439903, 763000416546, 763000439910, 763000629892, 763000629908, 00763000595906, 763000412685, 763000418151, 763000545499, 763000595906, 763000545505, 763000642235, 763000439866, 763000586195, 763000578640, 76300058618802, 76300058619501, 76300057864002, 76300054560402, 76300058619502
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.