Zap Surgical Systems: ZAP-X Radiosurgery System

Is my product affected?

ZAP-X Radiosurgery System

  • UPC 00860183001504

Check against the official notice below — it lists the full affected range.

What happened

Zap Surgical Systems has recalled the ZAP-X Radiosurgery System due to a software issue affecting patient setup. When there are large rotational deviations of 1.5 degrees or greater between digitally reconstructed radiographs and pre-delivery X-ray images, the system's transitional alignment algorithm may incorrectly calculate treatment table offset values, resulting in an incorrect position for the treatment isocenter. Eight units have been affected, including serial numbers ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, and ZUC21162, all running Treatment Delivery Software Version 1.8.53.

The affected devices were distributed worldwide, including in the U.S. states of Arizona, Maryland, Florida, and Colorado, as well as in Germany, Switzerland, and Japan. The recall record does not specify a remedy or recommended action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

Details from the source

Reason for recall: Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter. Distribution: Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN. Quantity: 8 Lot/code info: UDI-DI: 00860183001504, Serial Numbers, ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, ZUC21162, using Treatment Delivery Software Version: 1.8.53

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify