Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Reason for recall: Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. Distribution: US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX. Quantity: 14,379 kits total Lot/code info: 1. UDI-DI 10198459629334 (EA) 40198459629335 (CS) LOTS 26AMG895 2. UDI-DI 10198459545283 (EA) 40198459545284 (CS) LOTS 25LBJ366 26ABR365 3. UDI-DI 10195327471859 (EA) 40195327471850 (CS) LOTS 24EBA317 24EBV431 24IBD694 24JBQ783
10198459629334, 40198459629335, 10198459545283, 40198459545284, 10195327471859, 40195327471850
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.