Abbott: Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate,

Is my product affected?

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.

  • UPC 05414734500661
  • UPC 05414734509725

Check against the official notice below — it lists the full affected range.

What happened

Abbott is recalling the Merlin Patient Care System (PCS) Model 3650 Software Upgrade Kit due to a software defect. The software version 24.6.1 contains an anomaly where selecting an inappropriate shock energy value may cause the device to continue using a previously programmed energy level without alerting the user that the new selection was not accepted. This affects 4,733 programmers distributed across multiple U.S. states.

The recall involves programmers used to interrogate implantable cardioverter defibrillators (ICDs) that were used between January 22, 2020 and February 6, 2020. The source document does not specify a remedy or corrective action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.

Details from the source

Reason for recall: Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained. Distribution: U.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. OUS: None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to February 6, 2020. Quantity: Software Version 24.6.1 installed into 4733 programmers Lot/code info: Software Version: 24.6.1 Model: 3650, GTIN: 05414734500661 Model: 3330, GTIN: 05414734509725 For affected serial number of programmers see Attachment B.1 Programmers For models of ICDs interrogated by affected programmers see Attachment B.2 Devices.

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.