← Search recalls
FDA (Device enforcement)Recalled 2026-02-27class ii

Olympus Corporation of the Americas: Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU P

Hazard

Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

Details from the source

Reason for recall: Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong. Quantity: 408 units Lot/code info: Laser system: TFL-PLS; UDI: 00821925044111; Component Part: TFL-CPLU; UDI: 00821925044586; Serial #: MDUF220005, MDUF220005, MDUF220437, MDUF220437, MDUF220446, MDUF220446, MDUF190072, MDUF200324, MDUF210030, MDUF210204, MDUF210379, MDUF220001, MDUF220002, MDUF220003, MDUF220007, MDUF220008, MDUF220009, MDUF220010, MDUF220011, MDUF220012, MDUF220013, MDUF220014, MDUF220015, MDUF220016, MDUF220017, MDUF220018, MDUF220019, MDUF220020, MDUF220021, MDUF220022, MDUF220023, MDUF220024, MDUF220025, MDUF220026, MDUF220027, MDUF220028, MDUF220029, MDUF220030, MDUF220031, MDUF220032, MDUF220033, MDUF220034, MDUF220035, MDUF220036, MDUF220037, MDUF220038, MDUF220039, MDUF220040, MDUF220042, MDUF220043, MDUF220044, MDUF220045, MDUF220046, MDUF220047, MDUF220048, MDUF220049, MDUF220050, MDUF220051, MDUF220052, MDUF220053, MDUF220054, MDUF220055, MDUF220056, MDUF220057, MDUF220058, MDUF220060, MDUF220061, MDUF220062, MDUF220063, MDUF220064, MDUF220065, MDUF220066, MDUF220067, MDUF220069, MDUF220070, MDUF220071, MDUF220072, MDUF220074, MDUF220075, MDUF220076, MDUF220077, MDUF220078, MDUF220079, MDUF220080, MDUF220081, MDUF220082, MDUF220083, MDUF220084, MDUF220085, MDUF220086, MDUF220087, MDUF220088, MDUF220089, MDUF220090, MDUF220091, MDUF220092, MDUF220093, MDUF220094, MDUF220095, MDUF220096, MDUF220097, MDUF220098, MDUF220100, MDUF220101, MDUF220102, MDUF220103, MDUF220106, MDUF220109, MDUF220110, MDUF220111, MDUF220113, MDUF220114, MDUF220115, MDUF220116, MDUF220117, MDUF220119, MDUF220

UPC codes

00821925044111, 00821925044586

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify