Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Reason for recall: Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process. Distribution: Worldwide - US Nationwide and the countries of Canada. Quantity: 9,040 kits Lot/code info: 1. UDI-DI 10198459016257 (ea) 40198459016258 (case) 2. UDI-DI 10193489860078 (ea) 40193489860079 (case) ALL LOTS
10198459016257, 40198459016258, 10193489860078, 40193489860079