← Search recalls
FDA (Device enforcement)Recalled 2025-03-24class ii

Boston Scientific Corporation: IceSeed 1.5 CX S NEEDLE OUS, Cryoablation Needle, REF H7493967334100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or

Hazard

Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

Details from the source

Reason for recall: Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT. Distribution: Nationwide Quantity: 31 units Lot/code info: GTIN 00191506032685, Batch Numbers: 33996437, 33996438, 34295231, 34295232, 34392272, 34392273, 34392274, 34564921.

UPC codes

00191506032685

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.