An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
Reason for recall: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining. Distribution: Worldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions. Quantity: 8575 catheters Lot/code info: UDI/DI 08714729249504, Batch Numbers: 60442603, 60448206, 60463966, 60481407, 60491911, 60492607, 60443053, 60457140, 60467045, 60485945, 60492554, 60492608, 60443054, 60457142, 60467046, 60488392, 60492605, 60448186, 60463965, 60481406, 60488393, 60492606