← Search recalls
FDA (Device enforcement)Recalled 2025-03-17class ii

Abbott: Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080

Hazard

Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.

Details from the source

Reason for recall: Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism. Distribution: Worldwide - US Nationwide distribution. Quantity: 2683 Lot/code info: UDI-DI 00811806012926 Lots 9078966 9093895 9134623 9144113 9148110 9148579 9155868 9158302 9241591 9241593 10004059 10011461 10011462 10011463 10011465 10017790 10017792 10196106 10196119 10198795 10198801 10198889 10207472 10207503 10274454 10283818

UPC codes

00811806012926

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.