Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
Reason for recall: Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism. Distribution: Worldwide - US Nationwide distribution. Quantity: 4,850 Lot/code info: Models: 1. 9-TVLP4F90/060; UDI-DI 00811806011561 Lots 9114256 9125090 9157327 9211450 10005333 10080103 10092488 10092519 10212158 10248201 10265245 10333989 10334851 10353324 2. 9-TVLP4F90/080; UDI-DI 00811806011578 Lots 9206150 10040010 10068340 10227816 10341801
00811806011561, 00811806011578
Official record: FDA (Device enforcement) ↗