Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.
Reason for recall: Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment. Distribution: US Nationwide distribution including in the states of MO, MI, FL, PA, SC, OR, AZ, CO, ND, GA, NE, TX, IL, OH, VT, IN, TN, KY, IA, MT, NY, CA, UT, OK, SD, WA, AK, ME, NV, NM, NC, WV, LA, MN, ID, AL, VA, WI, NH, KS, MA, CT, RI, NJ, MD, MS, WY, DC, DE, HI, AR, ON. Quantity: 7,357 Lot/code info: REF/UDI-DI(GTIN)/Software Version: 139088-01/1.2.3.23; 169-137/10885403512544/2.4; 169-101/10885403512544/2.4; 169-102/10885403512544/2.4; 138902-01/2.4; 169-104/10885403512544/2.4; 169-105/10885403512544/2.4; 169-106/10885403512544/2.4; 138908-01/2.4; 169-107/10885403512544/2.4; 138906-01/2.4; 169-108/10885403512544/2.4; 169-110/10885403512544/2.4; 139091-01/2.4; 169-111/10885403512544/2.4; 138910-01/2.4; 169-112/10885403512544/2.4; 169-113/10885403512544/2.4; 138915-01/2.4; 169-23/10885403512568/2.4; 169-24/10885403512568/2.4; 138955-01/2.4; 169-25/10885403512568/2.4; 138924-01/2.4; 169-26/10885403512568/2.4; 138960-01/2.4; 169-27/10885403512568/2.4; 138961-01/2.4; 169-28/10885403512568/2.4; 138926-01/2.4; 138927-01/2.4; 169-29/10885403512568/2.4; 169-30/10885403512568/2.4; 169-31/10885403512568/2.4; 169-32/10885403512568/2.4; 169-33/10885403512568/2.4; 169-34/10885403512568/2.4; 169-139/10885403512568/2.4; 139125-01/2.4; 169-38/10885403512568/2.4; 138919-01/2.4; 169-39/10885403512568/2.4; 169-40/10885403512568/2.4; 169-41/10885403512568/2.4; 169-42/10885403512568/2.4; 138933-01/2.4; 169-140/10885403512568/1.2.3.23; 169-43/10885403512568/2.4; 169-44/10885403512568/2.4; 138959-01/2.4; 169-45/10885403512568/2.4; 169-46/10885403512568/2.4; 169-47/10885403512568/2.4; 169-49/10885403512568/2.4; 169-50/10885403512568/2.4; 138931-01/2.4; 169-52/10885403512568/2.4; 169-53/10885403512568/2.4; 138940-01/2.4; 169-54/10885403512568/2.4; 138952-01/2.4; 169-55/10885403512568/2.4; 169-56/
10885403512544, 10885403512568, 10885403517181, 10885403517167, 10885403512704
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.