During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury
Reason for recall: During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury Distribution: Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA, Australia, AUSTRIA, BANGLADESH, BARBADOS, BELGIUM, Bosnia and Herzegovina, BULGARIA, CANADA, CHILE, CROATIA, CZECH REPUBLIC, DENMARK, FRANCE, GREECE, Quantity: 746 Lot/code info: Model/UDI-DI; Software Version: 9735585/00763000631635, 9735586/00763000631765, 9735587/00763000631826; 3.1.4
00763000631635, 00763000631765, 00763000631826
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.