The metal ring at the end of the Guide Wire with Ruler Tube may detach from the main body of the instrument during use in surgery.
Reason for recall: The metal ring at the end of the Guide Wire with Ruler Tube may detach from the main body of the instrument during use in surgery. Distribution: US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NY, OH, TN, TX, WV. Quantity: 1588 units Lot/code info: Catalog number/REF: 2351-3080S; UDI-DI: 07613327361797; Lot #: KU155697, KU155698, KU155699, KU155700, KU155701, KU155702, KU155703, KU155704, KU155705, KU155706, KU155707, KU155708, KU155709, KU155710, KU155711, KU155712.
07613327361797
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.