← Search recalls
FDA (Device enforcement)Recalled 2023-04-05class ii

Beckman Coulter, Inc.: iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine

Hazard

Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.

Details from the source

Reason for recall: Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated. Distribution: Worldwide Distribution Quantity: 1068 units Lot/code info: a) iQ 200 Select, Part Numbers C10684, UDI/DI 15099590695484 and 700-3345, UD/DI 10837461001300; b) iQ 200 Elite, Part Numbers C10683, UDI/DI 15099590697129 and 700-3375, UDI/DI 10837461001355; c) iQ200 Select 1500, Part Number 700-3347, UD/DI 10837461001317; d) iQ200 Sprint, Part Number 700-3325, UDI/DI 10837461001256; All analyzers with Windows 10, Software Version 8.1;

UPC codes

15099590695484, 10837461001300, 15099590697129, 10837461001355, 10837461001317, 10837461001256

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify