There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Reason for recall: There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter. Distribution: Worldwide - US Nationwide distribution. Quantity: 41,739 units Lot/code info: All Serial Numbers. Product Code: CP150-2ES2, UDI-DI: 00732094192131; Product Code: CP150-2FR2, UDI-DI: 00732094192124; Product Code: CP150-2FR5, UDI-DI: 00732094192117; Product Code: CP150-2NL2, UDI-DI: 00732094192100; Product Code: CP150-2PT2, UDI-DI: 00732094192087; Product Code: CP150-2SV2, UDI-DI: 00732094192070; Product Code: CP150-3ENB, UDI-DI: 00732094179705; Product Code: CP150-3ESB, UDI-DI: 00732094179682; Product Code: CP150-4DE2, UDI-DI: 00732094192063; Product Code: CP150-4EN2, UDI-DI: 00732094179675; Product Code: CP150-4EN4, UDI-DI: 00732094179668; Product Code: CP150-4PT2, UDI-DI: 00732094192025; Product Code: CP150A-1EN2, UDI-DI: 00732094185157; Product Code: CP150A-1EN4, UDI-DI: 00732094185140; Product Code: CP150A-1EN66, UDI-DI: 00732094179644; Product Code: CP150A-1ENB, UDI-DI: 00732094177633; Product Code: CP150A-1ENBR, UDI-DI: 00732094322415; Product Code: CP150A-1ESB, UDI-DI: 00732094192230; Product Code: CP150A-1FRB, UDI-DI: 00732094179637; Product Code: CP150A-2DE2, UDI-DI: 00732094192018; Product Code: CP150A-2DE5, UDI-DI: 00732094192001; Product Code: CP150A-2EN2, UDI-DI: 00732094177626; Product Code: CP150A-2EN3, UDI-DI: 00732094179620; Product Code: CP150A-2EN4, UDI-DI: 00732094177619; Product Code: CP150A-2EN7, UDI-DI: 00732094191998; Product Code: CP150A-2ES2, UDI-DI: 00732094191981; Product Code: CP150A-2FR2, UDI-DI: 00732094191974; Product Code: CP150A-2FR5, UDI-DI: 00732094191967; Product Code: CP150A-2NL2, UDI-DI: 00732094191950; Product Cod
00732094192131, 00732094192124, 00732094192117, 00732094192100, 00732094192087, 00732094192070, 00732094179705, 00732094179682, 00732094192063, 00732094179675, 00732094179668, 00732094192025, 00732094185157, 00732094185140, 00732094179644, 00732094177633, 00732094322415, 00732094192230, 00732094179637, 00732094192018, 00732094192001, 00732094177626, 00732094179620, 00732094177619, 00732094191998, 00732094191981, 00732094191974, 00732094191967, 00732094191950, 00732094191943, 00732094191929, 00732094212938, 00732094179613, 00732094179606, 00732094184815, 00732094179590, 00732094179583, 00732094177602, 00732094213515, 00732094212952
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.