There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Reason for recall: There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter. Distribution: Worldwide - US Nationwide distribution. Quantity: 114,202 units Lot/code info: All Serial Numbers. Product Code: 7000-APM, UDI-DI: 00732094210613; Product Code: 75-HCA-CTB, UDI-DI: 00732094240597; Product Code: 75-HCA-MTB, UDI-DI: 00732094240603;
00732094210613, 00732094240597, 00732094240603
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.