← Search recalls
FDA (Device enforcement)Recalled 2023-10-05class ii

CareFusion 303, Inc.: MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software

Hazard

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

Details from the source

Reason for recall: If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal. Distribution: US Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ, IL, MA, NY, KY, UT, MD, WA, OH, AL. Quantity: 2679 Lot/code info: BD Pyxis MedBank Receiving Cabinet Software Software versions: 2.1.2.11, 2.2.1.8, 2.31.8, 2.3.2.1, 2.4.2.26 REF/UDI-DI: 1119-00/10885403512704; 1137-00; 1139-00/10885403512704; 1145-00/10885403517167; 1147-00/10885403517181; 139090-01; 169-100/10885403512568; 169-101/10885403512544; 169-102/10885403512544; 169-103/10885403512544; 169-104/10885403512544; 169-105/10885403512544; 169-106/10885403512544; 169-107/10885403512544; 169-108/10885403512544; 169-109/10885403512544; 169-110/10885403512544; 169-111/10885403512544; 169-112/10885403512544; 169-113/10885403512544; 169-132/10885403512568; 169-137/10885403512544; 169-138/10885403512568; 169-139/10885403512568; 169-153/10885403512568; 169-155/10885403512568; 169-156/10885403512568; 169-159/10885403512544; 169-160/10885403512544; 169-161/10885403512544; 169-162/10885403512544; 169-23/10885403512568; 169-24/10885403512568; 169-25/10885403512568; 169-26/10885403512568; 169-27/10885403512568; 169-28/10885403512568; 169-29/10885403512568; 169-30/10885403512568; 169-31/10885403512568; 169-32/10885403512568; 169-33/10885403512568; 169-34/10885403512568; 169-35/10885403512568; 169-36/10885403512568; 169-37/10885403512568; 169-38/10885403512568; 169-39/10885403512568; 169-40/10885403512568; 169-41/10885403512568; 169-42/10885403512568; 169-43/10885403512568; 169-44/10885403512568; 169-45/10885403512568; 169-46/10885403512568; 169-47/10885403512568; 169-48/10885403512568; 169-49/10885403512568; 169-50/10885403512568; 169-51/1088540351256

UPC codes

10885403512704, 10885403517167, 10885403517181, 10885403512568, 10885403512544

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.