Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Reason for recall: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded. Distribution: Worldwide distribution. Quantity: 3821 pumps Lot/code info: All serial numbers Medfusion Model 3500 pumps manufactured or serviced with these potentially affected Barrel Clamp Guides between Jul-2016 and Apr-2021. a) 3500, UDI/DI NONE, Serial Numbers: M02939, M03874, M09235, M09277, M09440, M09611, M09616, M10171, M10178, M10184, M10211, M10277, M10385, M10621, M11229, M12020, M12026, M12252, M12253, M12254, M13692, M14074, M14737, M15127, M15867, M15889, M16087, M16525, M17236, M17252, M18256, M18273, M18495, M18544, M18589, M18697, M18871, M19131, M21028, M21729, M21739, M22438, M24137, M24448, M28363, M28500, M28504, M30668, M30820, M32347, M36928, M39766; b) 3500-0600-00, UDI/DI 10610586041259, Serial Numbers: M107413, M107445, M107557, M107606, M107707, M107718, M110082, M110289, M110363, M111064, M73212, M-8833, M-8834, M-9284, M-9285, M-9286, M-9301, M-9302, M-9303; c) 3500-0600-01, UDI/DI 10610586041242, Serial Numbers: M100101, M102639, M102877, M103027, M103063, M103064, M103065, M103068, M103093, M103095, M103097, M103098, M103099, M103100, M103101, M103191, M103193, M103203, M103230, M103231, M103232, M103288, M103308, M103309, M103467, M103597, M103598, M103599, M103605, M103608, M103632, M103655, M103656, M103703, M103704, M103705, M103706, M103721, M103777, M103780, M103781, M103852, M103910, M103912, M103913, M103914, M103915, M103917, M103918, M103919, M103920, M103923, M103939, M103965, M103968, M104023, M104031, M104077, M104081, M104082, M104086, M104167, M104168, M104196, M104203, M104214, M104215, M104216, M10422
10610586041259, 10610586041242, 10610586041297, 30610586043288, 10610586043291, 10610586043017, 10610586030055, 10610586030031, 10610586030017, 10610586031175, 10610586032318
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.