ADVIA Chemistry Direct Bilirubin_2 assay-SMN: 10316610 (20 mL) & 10341114 (70 mL) - Product Usage: ADVIA DBIL_ 2 is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma on the ADVIA¿ Chemistry systems. Such measurements are useful in the screeni
Check against the official notice below — it lists the full affected range.
Siemens Healthcare Diagnostics is recalling ADVIA Chemistry Direct Bilirubin_2 assay kits in both 20 mL and 70 mL sizes. The assay is used for in vitro diagnostic testing to measure bilirubin levels. The recall affects approximately 1,901 units distributed in the United States and 37,349 units distributed internationally across Europe, Asia, Africa, Australia, and the Americas.
The reason for the recall is that the bilirubin assays may show a positive bias (higher readings than actual values) in patients taking eltrombopag therapy, a medication used to treat certain blood disorders. The affected lot codes span from February 2020 through July 2021. The recall document does not specify a remedy or recommended action for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
Reason for recall: Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy Distribution: Nationwide Foreign: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Kingdom, Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Azerbaijan, Baham Quantity: 1901 Us and 37349 OUS Lot/code info: ADVIA CH DBIL_2 (20 mL) 492068 (01)00630414513089(10)492068(17)20210728 ADVIA CH DBIL_2 (20 mL) 486501 (01)00630414513089(10)486501(17)20210628 ADVIA CH DBIL_2 (20 mL) 478164 (01)00630414513089(10)478164(17)20210328 ADVIA CH DBIL_2 (20 mL) 472667 (01)00630414513089(10)472667(17)20210228 ADVIA CH DBIL_2 (20 mL) 469068 (01)00630414513089(10)469068(17)20201228 ADVIA CH DBIL_2 (20 mL) 462448 (01)00630414513089(10)462448(17)20201028 ADVIA CH DBIL_2 (20 mL) 455476 (01)00630414513089(10)455476(17)20200728 ADVIA CH DBIL_2 (20 mL) 452841 (01)00630414513089(10)452841(17)20200628 ADVIA CH DBIL_2 (20 mL) 447337 (01)00630414513089(10)447337(17)20200528 ADVIA CH DBIL_2 (20 mL) 436518 (01)00630414513089(10)436518(17)20200228 ADVIA CH DBIL_2 (70 mL) 486509 (01)00630414513065(10)486509(17)20210628 ADVIA CH DBIL_2 (70 mL) 478169 (01)00630414513065(10)478169(17)20210328 ADVIA CH DBIL_2 (70 mL) 472672 (01)00630414513065(10)472672(17)20210228 ADVIA CH DBIL_2 (70 mL) 469073 (01)00630414513065(10)469073(17)20201228 ADVIA CH DBIL_2 (70 mL) 462453 (01)00630414513065(10)462453(17)20201028 ADVIA CH DBIL_2 (70 mL) 462459 (01)00630414513065(10)462459(17)20201028 ADVIA CH DBIL_2 (70 mL) 455481 (01)00630414513065(10)455481(17)20200728 ADVIA CH DBIL_2 (70 mL) 452846 (01)00630414513065(10)452846(17)20200628 ADVIA CH DBIL_2 (70 mL) 447351 (01)00630414513065(10)447351(17)20200528 ADVIA CH DBIL_2 (70 mL) 447357 (01)00630414513065(10)447357(17)20200528 ADVIA CH DBIL_2 (70 mL) 458975 (01)00630414513065(10)458975(
Official record: FDA (Device enforcement) ↗