There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
Reason for recall: There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low. Distribution: Worldwide distribution - US Nationwide. Quantity: 18,961 kits Lot/code info: ADVIA Centaur Cortisol (50T) UDI-DI: 00630414602943; Kit Lot Numbers: 49741344 07241346 88318346 29109348 and higher; ADVIA Centaur Cortisol 250T UDI-DI: 00630414602950; Kit Lot Numbers: 49742344 88319346 29108348 and higher; ADVIA Centaur Cortisol (REF) 250T UDI-DI: 00630414598659; Kit Lot Numbers: 75474344 11352346 and higher