Ion Beam Applications S.A.: IBA Proton Therapy System - PROTEUS 235

Hazard

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

Details from the source

Reason for recall: It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used. Distribution: Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan. Quantity: 4 units (1 US, 3 OUS) Lot/code info: PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)

UPC codes

05404013801138

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.