Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
Reason for recall: Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices. Distribution: International distribution in the country of Canada, Germany, Singapore, Australia, and India. Quantity: 55 units (All OUS) Lot/code info: Model/Catalog Number: SPL-SR, contains generator SPL-G; UDI: SPL-SR 00821925043824, SPL-G 00821925044203; Serial Numbers: CG5026, CG5027, CG5028, CG5029, CG5030, CG5031, CG5032, CG5033, CG5034, CG5035, CG5036, CG5037, CG5039, CG5040, CG5041, CG5042, CG5043, CG5045, CG5046, CG5047, CG5048, CG5049, CG5050, CG5051, CG5052, CG5053, CG5054, CG5055, CG5056, CG5057, CG5058, CG5059, CG5060, CG5061, CG5062, CG5063, CG5065, CG5066, CG5067, CG6037, CG6061, CG6069, CG6070, CG6071, CG6088, CG6094, CG6095, CG6096, CG6097, CG6098, CG6099, CG7000, CG7001, CG7002, CG7003;
00821925043824, 00821925044203
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.