← Search recalls
FDA (Device enforcement)Recalled 2019-04-05closedclass ii

Abbott Laboratories Inc. (St Jude Medical): TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)

Is my product affected?

TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)

  • UPC 07640157990194

Check against the official notice below — it lists the full affected range.

What happened

Abbott Laboratories Inc. (St Jude Medical) has recalled the TactiSys Quartz Equipment Model PN-004 400 with software version 1.7.0 due to a device log issue. In reported cases, the device's allocated disk space fills up, preventing the storage of new log data. This malfunction may cause contact force data to display intermittently during procedures. A total of 317 units have been distributed across more than 40 countries, including the United States, United Kingdom, and numerous locations in Europe, the Middle East, Asia, Africa, and the Americas.

The recalled devices are identified by specific serial numbers and UDI 07640157990194. The source document does not specify a remedy or corrective action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure.

Details from the source

Reason for recall: In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure. Distribution: US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA, PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, Quantity: 317 units Lot/code info: UDI: 07640157990194 Serial Numbers: 33019, 33059, 33350, 33354, 33357, 33360, 33361, 33363, 33390, 33392, 33393, 33405, 33417, 33421, 33444, 33449, 33451, 33457, 33458, 33461, 33491, 33492, 33509, 33013, 33015, 33020, 33040, 33075, 33091, 33097, 33160, 33197, 33384, 33385, 33387, 33389, 33414, 33433, 33388, 33435, 33481, 33154, 33157, 33162, 33163, 33278, 33092, 33159, 33437, 33412, 33014, 33027, 33028, 33029, 33038, 33049, 33057, 33082, 33096, 33192, 33222, 33268, 33379, 33383, 33397, 33398, 33399, 33400, 33422, 33423, 33478, 33320, 33352, 33395, 33351, 33069, 33381, 33455, 33064, 33093, 33313, 33337, 33161, 33426, 33083, 33321, 33420, 33051, 33155, 33196, 33256, 33353, 33460, 33217, 33427, 33073, 33150, 33076, 33148, 33434, 33396, 33050, 33333, 33333, 33336, 33336, 33095, 33143, 33143, 33112, 33113, 33266, 33267, 33269, 33270, 33271, 33272, 33273, 33274, 33275, 33276, 33283, 33287, 33288, 33291, 33325, 33356, 33359, 33362, 33404, 33459, 33016, 33098, 33099, 33100, 33101, 33102, 33103, 33104, 33105, 33106, 33107, 33108, 33110, 33111, 33127, 33130, 33131, 33265, 33018, 33024, 33025, 33087, 33152, 33156, 33220, 33230, 33231, 33254, 33279, 33281, 33315, 33416, 33428, 33438, 33439, 33442, 33443, 33463, 33330, 33386, 33228, 33033, 33048, 33149, 33259, 33282, 33464, 33022, 33031, 33035, 33037, 33044, 33045, 33046, 33055, 33066, 33067, 33068, 33078, 33080, 33085, 33086, 33088, 33094, 33129, 33137, 33138, 33140, 33144, 33158, 33195, 33218, 33221, 33229, 33253, 33257, 33262, 33280, 3

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify