Brilliance BigBore Radiology CT - 728244 Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and e
Check against the official notice below — it lists the full affected range.
Philips Medical Systems is recalling 67 Brilliance BigBore Oncology/Radiology CT systems with software version 4.2.0 distributed worldwide, including locations in the United States and internationally. The systems have a defect affecting 4D CT pulmonary imaging: when users perform retrospective respiratory gated 4D CT scans using the Bellows pulmonary gating device, the CT images may be reconstructed at only a single phase while the image labels incorrectly indicate they represent multiple requested phases ranging from 0% to 90%.
The recall was triggered by customer complaints about this imaging issue. The affected systems can be identified by their specific serial numbers listed in the recall. The source document does not specify what affected consumers should do in response to this recall.
Written by software from the official notice below. Not reviewed by a lawyer.
Philips identified a 4D CT pulmonary phase issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software version through customer complaints. When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).
Reason for recall: Philips identified a 4D CT pulmonary phase issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software version through customer complaints. When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%). Distribution: Worldwide Distribution - US to AR, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, TX & WI; Internationally to the following countries: Australia, Austria, Canada, China, Czech Republic, Germany, India, Israel, Jordan, Korea, Mexico, New Cldonia, Romania, Russian Federation, Turkey, & United Kingdom. Quantity: 67 Lot/code info: System Serial Number - 76001, 76005, 76007, 76009, 76013, 76014, 76017, 76018, 76019, System Serial Number - Unique Device Identifier 76058 (01)00884838059450(21)76058 76033 (01)00884838059450(21)76033 76048 (01)00884838059450(21)76048 76059 (01)00884838059450(21)76059 76015 (01)00884838059450(21)76015 76063 (01)00884838059450(21)76063 76066 (01)00884838059450(21)76066 76049 (01)00884838059450(21)76049 76040 (01)00884838059450(21)76040 76032 (01)00884838059450(21)76032 76061 (01)00884838059450(21)76061 76031 (01)00884838059450(21)76031 76006 (01)00884838059450(21)76006 76041 (01)00884838059450(21)76041 76035 (01)00884838059450(21)76035 76038 (01)00884838059450(21)76038 76065 (01)00884838059450(21)76065 76064 (01)00884838059450(21)76064 76053 (01)00884838059450(21)76053 76050 (01)00884838059450(21)76050 76060 (01)00884838059450(21)76060 76036 (01)00884838059450(21)76036 76056 (01)00884838059450(21)76056 76055 (01)00884838059450(21)76055 76037 (01)00884838059450(21)76037 76051 (01)00884838059450(21)76051 76044 (01)00884838059450(21)76044 76039 (01)00884838059450(21)76039 76012 (01)00884838059450(21)76012 76026 (01)00884838059450(21)76026 760016 (01)00884838059450(21)760016 760001 (01)00884838059450(21)760001 760005 (01)00884838059450(21)760005 760004 (01)00884838059450(21)760004 760003 (01)00884838059450(21)760003 760007 (01)00884838059450(21)760007 76072 (01)00884838059450(21)76072 76062 (01)00884838059450(21)76062 76043 (01)00884838059450(21)76043 760013 (01)00884838059450(21
Official record: FDA (Device enforcement) ↗