Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.
Reason for recall: Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide. Distribution: US: TX, NE, and AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, FRANCE, GERMANY, ITALY, SPAIN, SWEDEN, SWITZERLAND, UK Quantity: 47 units Lot/code info: UDI/DI 00889797004008, Lot Numbers: 2656950, 2762126, 2796094
00889797004008