What happened

Becton Dickinson & Company is recalling BD Connecta Stopcock Without Extension Tube (Catalog No. 394605) distributed worldwide, including across numerous U.S. states and countries in Asia, Europe, and other regions. The stopcock may leak at the housing component, which could result in treatment delays or interruptions, exposure to infusate and biohazardous material, under-dosing, contamination, and/or air ingress into the line. The recall affects specific lot numbers with expiration dates ranging from January 31, 2024, through April 30, 2024.

The recall notice does not specify a remedy or recommended action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.

Details from the source

Reason for recall: May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress. Distribution: Worldwide distribution - US Nationwide distribution in the states of AK, AR, AZ, CA, CO, FL, GA, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WV and the countries of Belgium, China, India, Indonesia, Japan, Korea, Nepal, Philippines, Singapore, Sri Lanka, Thailand, Vietnam. Quantity: N/A Lot/code info: GTIN: 382903946051 Lot Numbers/Exp.Date: 1133181 4/30/2024 1117434 3/31/2024 1110835 3/31/2024 1119671 3/31/2024 1098353 3/31/2024 1098351 3/31/2024 1098350 3/31/2024 1110839 3/31/2024 1110842 3/31/2024 1067265 2/29/2024 1067268 2/29/2024 1067274 2/29/2024 1063372 2/29/2024 1067273 2/29/2024 1067264 2/29/2024 1034587 1/31/2024 1040899 1/31/2024 1034590 1/31/2024 1034594 1/31/2024 1040902 1/31/2024 1040891 1/31/2024 1034597 1/31/2024 1060806 2/29/2024

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify