The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
Reason for recall: The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). Distribution: Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand Quantity: 4007 units Lot/code info: a) Item Number 314-13-32, UDI/DI 10885862201003, Serial Numbers: 3864654, 3864659, 4140193, 4155690, 4257179, 4250984, 4155679, 4140217, 4155678, 3864658, 4108679, 4155704, 4257186, 4257198, 3926702, 4108676, 4257194, 4108680, 3926703, 4155692, 3864663, 3864662, 3864653, 3926698, 4257174, 4108685, 4269883, 4155683, 4257173, 4108681, 4269885, 4269868, 4155685, 4155677, 4269870, 4155700, 4108682, 4108683, 4269898, 4155697, 4140198, 4257180, 4257196, 4269876, 4968395, 4684348, 4684355, 4684340, 4684354, 4808076, 4808075, 4968389, 4757115, 4684353, 4757123, 4808079, 4968386, 5067509, 4808072, 4808060, 4269872, 4257193, 3864650, 4250985, 4108678, 4108684, 4257185, 4250982, 4087951, 4269882, 4140207, 4250976, 4257187, 4269900, 3864657, 4250992, 3926700, 4269873, 4968393, 5088349, 5304460, 5304457, 4808070, 5304512, 4684342, 5304461, 5304479, 4684347, 5304478, 5519088, 5088351, 4684344, 5304475, 5519087, 4968399, 5304477, 5304485, 4684365, 4684350, 5519079, 4757113, 4757124, 5519094, 5304459, 5304487, 5519083, 4684359, 5304469, 5304486, 5563435, 5563436, 4257167, 4155705, 4140205, 4257190, 4155682, 4155715, 4250979, 4250980, 4155703, 3926701, 4269884, 4140192, 4250968, 4257175, 4155706, 4269890, 4257176, 4250983, 4155686, 4140212, 3864664, 4140200, 4269897, 4257188, 4257172, 5086602, 5304480, 4808068, 5304498, 5304470, 5304484, 5116203, 5088348, 5088350, 4808078, 4684361, 4684362, 5519093, 4684352, 5519084, 5304488, 5304465, 5519092, 4968387, 5304503, 5304500, 5304515, 4968385, 4968
10885862201003, 10885862201010, 10885862201027
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.