Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.
Reason for recall: Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site. Distribution: US and Australia, Austria, Belgium, France, Germany, Greece, Italy, Netherlands, Spain, Sweden, Switzerland Quantity: 601 needles Lot/code info: Lot number 837838839 appears on the labels for REF 911933, REF 911761, and REF 912465. UDI numbers: REF 911933 - (01)0 7340048 30034 6, (17)250401 (10) 837838839 (240) 911933; REF 911761 - (01)0 7340048 30800 7, (17)250401 (10) 837838839 (240) 911761; REF 912465 - (01)0 7340048 30801 4, (17)250401 (10) 837838839 (240) 912465.
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.