Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Reason for recall: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. Distribution: Worldwide distribution - US Nationwide. Quantity: 551 units Lot/code info: 1) DYNJ0185290G, UDI-DI: 10193489378054(each), 40193489378055(case), Lot Number: 22LBQ336; 2) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 24BBN429; 3) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23KBI089; 4) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23HBG596; 5) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBL644; 6) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBG152; 7) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23EBH550; 8) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23CBQ546; 9) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22NBF780; 10) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22LBC300; 11) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBP453; 12) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBM263; 13) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBF686; 14) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22CBY841; 15) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBR457; 16) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBA253; 17) DYNJ905567A, UDI-DI: 10193489348927(each), 40193
10193489378054, 40193489378055, 10193489348927, 40193489348928
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.