← Search recalls
FDA (Device enforcement)Recalled 2024-02-20class i

Boston Scientific Corporation: Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010

Hazard

An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.

Details from the source

Reason for recall: An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures. Distribution: US Quantity: 985 units Lot/code info: a) M0013972001010, UDI/DI 00191506039332, ALL LOT CODES b) M0013972101010, UDI/DI 00191506043124, ALL LOT CODES

UPC codes

00191506039332, 00191506043124

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify