Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only. Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids. This analyzer is optimized for high throu
Check against the official notice below — it lists the full affected range.
Roche Diagnostics Operations, Inc. is recalling 72 units of Cobas 8000 modular analyzer series software versions 03-01, 02, 03, and 04 that impact the cobas e 602 module. The issue involves a software defect where Diluent Multiassay may be incorrectly used for autodilution of certain assays instead of Diluent Universal. This can produce incorrect test results only when specific conditions are met simultaneously: more than 2000 total reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluents are involved. The affected devices were distributed nationwide in the USA.
The recall affects 72 specific units with part number 05990378001 and various serial numbers listed in the recall notice. The record does not specify a remedy for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met simultaneously: More than a total of 2000 reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluent
Reason for recall: When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met simultaneously: More than a total of 2000 reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluent Distribution: USA Nationwide Distribution Quantity: 72 units Lot/code info: Part Number 05990378001 Serial Numbers 1017-06, 1021-03, 1021-05, 1142-09, 1145-10, 1140-01, 1125-07, 1128-02, 1146-01, 1146-02, 1146-03, 1171-10, 1169-06, 1169-08, 1169-09, 1172-09, 1156-01, 1156-02, 1156-09, 1160-04, 1160-06, 1016-05, 1017-01, 1139-08, 1139-10, 1140-07, 1140-08, 1140-10, 1156-06, 1156-07, 1158-09, 1160-01, 1160-05, 1173-01, 1139-06, 1139-07, 2374-11, 1127-08, 1138-08, 1138-10, 1139-01, 1156-08, 1156-10, 1173-10, 1156-05, 1158-10, 1139-09, 1131-08, 1132-01, 1132-02, 1171-07, 1171-08, 1171-09, 1147-05, 1149-01, 1124-05, 1124-06, 1132-03, 1132-04, 1147-05, 1149-01, 1128-05, 1128-06, 1139-02, 1139-03, 1139-04, 1139-05, 1146-05, 1146-06, 1147-04, 1156-04, 1132-05, 1132-10, 1137-02, 1147-05, 1149-01, 1128-05, 1128-06, 1149-07, 1149-08, and 1167-02.
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.