Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.
Reason for recall: Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication. Distribution: Worldwide distribution - US Nationwide and the country of Bahamas. Quantity: 926,227 units Lot/code info: 1) DYNDTB0529, UDI-DI: 10197344105663(each), 20197344105660(case), Lot Number: All Lot Numbers; 2) DYNDTB1029, UDI-DI: 10197344076956(each), 20197344076953(case), Lot Number: All Lot Numbers; 3) DYNDTB1516, UDI-DI: 10197344105687(each), 20197344105684(case), Lot Number: All Lot Numbers; 4) DYNDTB1530, UDI-DI: 10197344105694(each), 20197344105691(case), Lot Number: All Lot Numbers; 5) DYNDTB1539, UDI-DI: 10197344008254(each), 20197344008251(case), Lot Number: All Lot Numbers; 6) DYNDTB1545, UDI-DI: 10197344098644(each), 20197344098641(case), Lot Number: All Lot Numbers; 7) DYNDTB2045, UDI-DI: 10197344077007(each), 20197344077004(case), Lot Number: All Lot Numbers; 8) DYNDTB0512, UDI-DI: 10197344105656(each), 20197344105653(case), Lot Number: All Lot Numbers; 9) DYNDTB0555, UDI-DI: 10197344105670(each), 20197344105677(case), Lot Number: All Lot Numbers; 10) DYNDTB1022, UDI-DI: 10197344076949(each), 20197344076946(case), Lot Number: All Lot Numbers; 11) DYNDTB1512, UDI-DI: 10197344098613(each), 20197344098610(case), Lot Number: All Lot Numbers; 12) DYNDTB1529, UDI-DI: 10197344098620(each), 20197344098627(case), Lot Number: All Lot Numbers; 13) DYNDTB1537, UDI-DI: 10197344008247(each), 20197344008244(case), Lot Number: All Lot Numbers; 14) DYNDTB1540, UDI-DI: 10197344098637(each), 20197344098634(case), Lot Number: All Lot Numbers; 15) DYNDTB2029, UDI-DI: 10197344076994(each), 20197344076991(case), Lot Number: All Lot Numbers; 16) DYNDTB5077, UDI-DI: 10197344098651(each), 20197344
10197344105663, 20197344105660, 10197344076956, 20197344076953, 10197344105687, 20197344105684, 10197344105694, 20197344105691, 10197344008254, 20197344008251, 10197344098644, 20197344098641, 10197344077007, 20197344077004, 10197344105656, 20197344105653, 10197344105670, 20197344105677, 10197344076949, 20197344076946, 10197344098613, 20197344098610, 10197344098620, 20197344098627, 10197344008247, 20197344008244, 10197344098637, 20197344098634, 10197344076994, 20197344076991, 10197344098651, 20197344098658, 10197344076963, 20197344076960, 10197344105700, 20197344105707, 10197344105717, 20197344105714, 10197344076932, 20197344076939
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.