FDA (Device enforcement)Recalled 2023-03-24class i

Covidien: Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH

Hazard

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

Details from the source

Reason for recall: A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding. Distribution: Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA, TX, WI, CT, MA, WV, CO, VT, NJ, SD, AR, AL, MS, LA, DE, MT, WY, DC, UT, NV, NH, RI, ID, NM, ME and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bolivia, Plurinational State O Quantity: 599,187 Lot/code info: REF/UDI-DI/Lot: 10CN10H/10884521172371/21C0402JZX, 21D0461JZX, 21G0457JZX, 21I0293JZX, 21I0294JZX, 21I0552JZX, 21J1435JZX, 21J1436JZX, 21K0275JZX, 21K0276JZX, 21K0277JZX, 21K0278JZX; 10CN10R/10884521205475/202107218X, 202107296X, 202108128X, 202108285X, 21H0792JZX, 21H0793JZX, 21H0794JZX, 22I0596JZX; 10UN10A/20884521175690/21D0463JZX; 10UN10H/20884521176291/21E0661JZX, 21G0473JZX; 10UN10R/10884521205543/202105307X, 202105308X, 202106148X, 202209335X; 4CN65A/10884521172456/21C0432JZX, 21I0369JZX; 4CN65H/20884521176307/21C0358JZX, 21C0359JZX, 21C0360JZX, 21C0361JZX, 21D0440JZX, 21D0441JZX, 21E0512JZX, 21E0513JZX, 21E0514JZX, 21E0515JZX, 21E0516JZX, 21E0517JZX, 21E0518JZX, 21E0519JZX, 21E0520JZX, 21E0521JZX, 21E0522JZX, 21E0523JZX, 21E0524JZX, 21E0525JZX, 21E0526JZX, 21E0527JZX, 21E1773JZX, 21E1774JZX, 21E1775JZX, 21E1776JZX, 21E1777JZX, 21E1778JZX, 21E1779JZX, 21E1780JZX, 21E1781JZX, 21E1782JZX, 21E1783JZX, 21H1324JZX, 21J0181JZX, 21J1094JZX, 21K0237JZX, 21K0239JZX, 21K0639JZX, 21K0643JZX, 21K0645JZX, 22C1465JZX, 22C1466JZX, 22C1467JZX, 22D0798JZX, 22D0799JZX, 22D0800JZX, 22D0801JZX, 22D0891JZX, 22D0892JZX, 22D0893JZX, 22D0897JZX, 22D0898JZX, 22D0899JZX, 22D0900JZX, 22D0901JZX, 22D0909JZX, 22F0918JZX, 22F0919JZX, 22F0920JZX, 22F0921JZX, 22F0922JZX, 22F0923JZX, 22F0924JZX, 22F0925JZX, 22F0926JZX, 22F0927JZX, 22F0928JZX, 22F0929JZX, 22J0073JZX, 22J0074JZX, 22J0075JZX, 22J0076JZX, 22J0077JZX, 22J0080JZX, 22J0081JZX, 22J0083JZX; 4CN65R/10884521205024/202105082X, 202105310X, 2021062

UPC codes

10884521172371, 10884521205475, 20884521175690, 20884521176291, 10884521205543, 10884521172456, 20884521176307, 10884521205024, 20884521175706, 20884521176239, 10884521205482, 10884521172463, 20884521176314, 10884521205420, 20884521175713, 20884521176246, 10884521205499, 10884521172470, 20884521176321, 10884521205437, 20884521176253, 10884521205505, 10884521172487, 20884521176338, 10884521205444, 20884521175737, 20884521176260, 10884521205512, 10884521172494, 20884521176345, 10884521205451, 20884521175744, 20884521176277, 10884521205529, 20884521176352, 10884521205468, 20884521176284, 10884521205536

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify