A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.
Reason for recall: A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding. Distribution: Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA, TX, WI, CT, MA, WV, CO, VT, NJ, SD, AR, AL, MS, LA, DE, MT, WY, DC, UT, NV, NH, RI, ID, NM, ME and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bolivia, Plurinational State O Quantity: 599,187 Lot/code info: REF/UDI-DI/Lot: 10CN10H/10884521172371/21C0402JZX, 21D0461JZX, 21G0457JZX, 21I0293JZX, 21I0294JZX, 21I0552JZX, 21J1435JZX, 21J1436JZX, 21K0275JZX, 21K0276JZX, 21K0277JZX, 21K0278JZX; 10CN10R/10884521205475/202107218X, 202107296X, 202108128X, 202108285X, 21H0792JZX, 21H0793JZX, 21H0794JZX, 22I0596JZX; 10UN10A/20884521175690/21D0463JZX; 10UN10H/20884521176291/21E0661JZX, 21G0473JZX; 10UN10R/10884521205543/202105307X, 202105308X, 202106148X, 202209335X; 4CN65A/10884521172456/21C0432JZX, 21I0369JZX; 4CN65H/20884521176307/21C0358JZX, 21C0359JZX, 21C0360JZX, 21C0361JZX, 21D0440JZX, 21D0441JZX, 21E0512JZX, 21E0513JZX, 21E0514JZX, 21E0515JZX, 21E0516JZX, 21E0517JZX, 21E0518JZX, 21E0519JZX, 21E0520JZX, 21E0521JZX, 21E0522JZX, 21E0523JZX, 21E0524JZX, 21E0525JZX, 21E0526JZX, 21E0527JZX, 21E1773JZX, 21E1774JZX, 21E1775JZX, 21E1776JZX, 21E1777JZX, 21E1778JZX, 21E1779JZX, 21E1780JZX, 21E1781JZX, 21E1782JZX, 21E1783JZX, 21H1324JZX, 21J0181JZX, 21J1094JZX, 21K0237JZX, 21K0239JZX, 21K0639JZX, 21K0643JZX, 21K0645JZX, 22C1465JZX, 22C1466JZX, 22C1467JZX, 22D0798JZX, 22D0799JZX, 22D0800JZX, 22D0801JZX, 22D0891JZX, 22D0892JZX, 22D0893JZX, 22D0897JZX, 22D0898JZX, 22D0899JZX, 22D0900JZX, 22D0901JZX, 22D0909JZX, 22F0918JZX, 22F0919JZX, 22F0920JZX, 22F0921JZX, 22F0922JZX, 22F0923JZX, 22F0924JZX, 22F0925JZX, 22F0926JZX, 22F0927JZX, 22F0928JZX, 22F0929JZX, 22J0073JZX, 22J0074JZX, 22J0075JZX, 22J0076JZX, 22J0077JZX, 22J0080JZX, 22J0081JZX, 22J0083JZX; 4CN65R/10884521205024/202105082X, 202105310X, 2021062
10884521172371, 10884521205475, 20884521175690, 20884521176291, 10884521205543, 10884521172456, 20884521176307, 10884521205024, 20884521175706, 20884521176239, 10884521205482, 10884521172463, 20884521176314, 10884521205420, 20884521175713, 20884521176246, 10884521205499, 10884521172470, 20884521176321, 10884521205437, 20884521176253, 10884521205505, 10884521172487, 20884521176338, 10884521205444, 20884521175737, 20884521176260, 10884521205512, 10884521172494, 20884521176345, 10884521205451, 20884521175744, 20884521176277, 10884521205529, 20884521176352, 10884521205468, 20884521176284, 10884521205536
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.