Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
Reason for recall: Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality. Distribution: US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA. Quantity: 47 Lot/code info: UDI: 00886874119747, Serial: SQ0326, SQ0336, SQ0341, SQ0345, SQ0346, SQ0356, SQ0367, SQ0368, SQ0375, SQ0377, SQ0379, SQ0380, SQ0386, SQ0387, SQ0389, SQ0394, SQ0395, SQ0398, SQ0401, SQ0404, SQ0412, SQ0413, SQ0415, SQ0419, SQ0420, SQ0422, SQ0425, SQ0427, SQ0428, SQ0432, SQ0433, SQ0436, SQ0443, SQ0447, SQ0448, SQ0451, SQ0454, SQ0457, SQ0458, SQ0462, SQ0464, SQ0469, SQ0470, SQ0662, SQ0727, SQ0740, SQ0761
00886874119747
Official record: FDA (Device enforcement) ↗