Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Reason for recall: Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active. Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Quantity: 802 units (45 US, 757 OUS) Lot/code info: System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); UDI-DIs: (1) 884838085282, (2) 884838099210, (3) 884838116733, (4)N/A; Serial Numbers: All;
884838085282, 884838099210, 884838116733