Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
Reason for recall: Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port. Distribution: US Nationwide. Belgium, Canada, Costa Rica, Thailand. Quantity: 1412 units Lot/code info: UDI-DI: (01)00699753450134; Lot Number: 2M10
00699753450134
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.