Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications
Reason for recall: Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications Distribution: US Nationwide distribution. Quantity: 1225 units (packs of 10) Lot/code info: GTIN: 07613327606157 Lot Numbers; 2022111803 2022112901 2022121911 2022121912 2022122601 2022122602
07613327606157
Official record: FDA (Device enforcement) ↗