← Search recalls
FDA (Device enforcement)Recalled 2024-02-09class ii

Preat Corp: Preat Nobel Active/Conical-compatible RP Titanium Blank, REF 9000967

Hazard

Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

Details from the source

Reason for recall: Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant. Distribution: U.S. Nationwide distribution in the states of AZ, OH and PA. Quantity: 561 units Lot/code info: Model Number: 9000967 UDI-DI code: 00842092165867 Lot Numbers: 235650, 234294, 233667, 233885, 233439, and 231489

UPC codes

00842092165867

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.