← Search recalls
FDA (Device enforcement)Recalled 2026-01-07class ii

Philips North America Llc: Spectral CT; Product Code (REF): 728333;

Hazard

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Details from the source

Reason for recall: As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled. Distribution: Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, J Quantity: 283 units Lot/code info: Product Code (REF): 728333; UDI-DI: 00884838101111; Serial Numbers: 10055, 10075, 10086, 10096, 10106, 10116, 10126, 10136, 10146, 10156, 10166, 10176, 10186, 10196, 10206, 10217, 10227, 10237, 10247, 10257, 10267, 10277, 10289, 10302, 10314, 12009, 12019, 12029, 396013, 10056, 10076, 10087, 10097, 10107, 10117, 10127, 10137, 10147, 10157, 10167, 10177, 10187, 10197, 10207, 10218, 10228, 10238, 10248, 10258, 10268, 10278, 10290, 10303, 12000, 12010, 12020, 12030, 398003, 10057, 10077, 10088, 10098, 10108, 10118, 10128, 10138, 10148, 10158, 10168, 10178, 10188, 10198, 10208, 10219, 10229, 10239, 10249, 10259, 10269, 10279, 10291, 10304, 12001, 12011, 12021, 12031, 398004, 10058, 10078, 10089, 10099, 10109, 10119, 10129, 10139, 10149, 10159, 10169, 10179, 10189, 10199, 10209, 10220, 10230, 10240, 10250, 10260, 10270, 10280, 10292, 10305, 12002, 12012, 12022, 396005, 10059, 10079, 10090, 10100, 10110, 10120, 10130, 10140, 10150, 10160, 10170, 10180, 10190, 10200, 10210, 10221, 10231, 10241, 10251, 10261, 10271, 10281, 10293, 10306, 12003, 12013, 12023, 396006, 10060, 10080, 10091, 10101, 10111, 10121, 10131, 10141, 10151, 10161, 10171, 10181, 10191, 10201, 10211, 10222, 10232, 10242, 10252, 10262, 10272, 10284, 10294, 10307, 12004, 12014, 12024, 396008, 10061, 10081, 10092, 10102, 10112, 10122, 10132, 10142, 10152, 10162, 10172, 10182, 10192, 10202, 10212, 10223, 10233, 10243, 10253, 10263, 10273, 10285, 10298, 10308, 12005, 12015, 12025, 396009, 10062, 10082, 10093, 10103, 1011

UPC codes

00884838101111

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify