Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures.
Check against the official notice below — it lists the full affected range.
Medtronic is recalling certain Bio-Console 560 Extracorporeal Blood Pumping Console devices used during heart surgery procedures. An electrical component within the system controller module of these devices may fail, causing the pump to stop working during use. Problems reported include the pump stopping unexpectedly, the user interface screen going blank, and the device smoking or producing a burning odor.
The recall affects 387 devices worldwide, including 93 in the United States, distributed across multiple countries. The devices span several model numbers (560BC, 560BC1, 560BCS, 560BCS1) and include specific serial numbers identified in the recall details. The source document does not specify what action affected consumers or healthcare facilities should take in response to this recall.
Written by software from the official notice below. Not reviewed by a lawyer.
An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.
Reason for recall: An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use. Distribution: Worldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico, Morocco, New Zealand, Russian Federation, Spain, Taiwan, Thailand, United Kingdom, and Viet Nam Quantity: 387 devices (93 US) Lot/code info: 560BC: GTIN 00885074539386, Serial Numbers: 560B100094, 560B100212, 560B100244, 560B100257, 560B100274, 560B100283, 560B100418, 560B100439, 560B100482 560BC1: GTIN 00613994450326, Serial Numbers: 560B100901, 560B101368 560BCS: GTIN 00613994258236, Serial Numbers: 560B100369, 560B100376, 560B100472, 560B100495, 560B100496 560BCS1: GTIN: 00613994450463, Serial Numbers: 560B100191, 560B100690, 560B100783, 560B100806, 560B100820, 560B100871, 560B100896, 560B101062, 560B101187, 560B101318, 560B101344, 560B101470, 560B101499, 560B101504, 560B101519, 560B101593, 560B101526 GTIN 00613994526052, Serial Numbers: 560B102610, 560B102725 GTIN 00643169201033, Serial Numbers: 560B101838 GTIN 00643169201651, Serial Numbers: 560B101834 GTIN 00643169315822, Serial Numbers: 560B100127, 560B101694, 560B101700, 560B101805, 560B101976, 560B101996, 560B102080, 560B102118, 560B102185, 560B102194, 560B102204, 560B102210, 560B102215, 560B102265, 560B102336, 560B102426, 560B102493, 560B102518, 560B102547, 560B102559, 560B102598, 560B102761, 560B102762, 560B102763, 560B102764, 560B102765, 560B102766 GTIN 00763000135447, Serial Numbers: 560B102711, 560B102717, 560B102718, 560B102720, 560B102721, 560B102722, 560B102723, 560B102724, 560B102726, 560B102727, 560B102730, 560B102731, 560B102732, 560B102733, 560B102734, 560B102735, 560B102736, 560B102737, 560B102738, 560B102739, 560B102740, 560B102741, 560B102742, 560B102743, 560B102745, 560B102746, 560B102747, 560B102748, 560B102750, 560B102751, 560B102752, 56
Official record: FDA (Device enforcement) ↗