Product Name: GCX Channel Mounting Accessory Model/Catalog Number: H8-300-40-A
Check against the official notice below — it lists the full affected range.
Skytron's GCX Channel Mounting Accessory Model H8-300-40-A has been recalled because the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when used with the VST clamp as a monitor mount. A total of 731 units have been affected and distributed worldwide, including throughout the US, Canada, and Japan.
All units of this model are included in the recall. Ergon 3 booms manufactured after September 2016 can be identified by UDI 10841736100889, while units manufactured prior to September 2016 do not have a UDI. The source document does not specify a remedy or what action affected consumers should take.
Written by software from the official notice below. Not reviewed by a lawyer.
Under certain circumstances, the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when the GCX Channel Mounting Accessory was used with the VST clamp as a monitor mount on the boom.
Reason for recall: Under certain circumstances, the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when the GCX Channel Mounting Accessory was used with the VST clamp as a monitor mount on the boom. Distribution: Worldwide distribution - US Nationwide and the countries of Canada and Japan. Quantity: 731 Lot/code info: Lot Code: All units of model H8-300-40-A are affected. Code Info: GCX mounting bracket does not have UDI. Ergon 3 booms after September 2016 have UDI. DI of units manufactured after September 2016 is 10841736100889. Units manufactured prior to September 2016 do not have UDI.
Official record: FDA (Device enforcement) ↗