GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Reason for recall: GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls. Distribution: Worldwide distribution. Quantity: 21 units Lot/code info: System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 650493REV2, UDI To be provided, Mfg. Lot/Serial Number REVVX1900036CN; System ID Number 864512APEX1, UDI (01)00195278460271(11)240500(21)REV2X2400059CN, Mfg. Lot/Serial Number REV2X2400059CN; System ID Number 832325REV, UDI (01)00195278460271(11)241000(21)REV2X2400134CN, Mfg. Lot/Serial Number REV2X2400134CN; System ID Number 281MWREVCT, UDI To be provided, Mfg. Lot/Serial Number REVVX1900008CN; System ID Number 713792ACT39, UDI To be provided, Mfg. Lot/Serial Number REVV82100006CN; System ID Number 206320REV, UDI To be provided, Mfg. Lot/Serial Number 00000441459CN5; System ID Number 206320REV2, UDI To be provided, Mfg. Lot/Serial Number REVV82200025CN; System ID Number 082421190701, UDI (01)00195278633149(11)241100(21)CBFCG2400026HM, Mfg. Lot/Serial Number CBFCG2400026HM; System ID Number DK1044CT12, UDI (01)00840682146616(11)200200(21)REV2A2000001CN, Mfg. Lot/Serial Number REV2A2000001CN; System ID Number DK1547CT02, UDI (01)00840682146616(11)210100(21)REV2A2100002CN, Mfg. Lot/Serial Number REV2A2100002CN; System ID Number 11525316, UDI To be provided, Mfg. Lot/Serial Number REV2X2400082CN; System ID Number 259056CT01, UDI (01)00195278460271(11)230800(21)REV2X2300051CN, Mfg. Lot/Serial Number REV2X2300051CN; System ID Number 11437190, UDI To be provided, Mfg. Lot/Serial Number CBGLG2400001HM; System ID Number 0850211094, UDI (01)00195278
00195278460271, 082421190701, 00195278633149, 00840682146616, 00840682123471, 00840682118552
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.