Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
Reason for recall: Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results. Distribution: Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Bangladesh, Bolivia, Brazil, Canada, China, Croatia, Czech Republic, Denmark, Egypt, El Salvador, Georgia, Greece, Hong Kong, India, Kazakhstan, Mexico, Namibia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Taiwan, Thailand, Turkey, Quantity: 64,009 kits Lot/code info: Lot numbers: 338555, exp. 31 JUL 2024; 338556, exp. 31 AUG 2024; 338734, exp. 30 SEP 2024; and 338958, exp. 31 OCT 2024; UDI-DI 15099590225742.
15099590225742
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.