FDA (Device enforcement)Recalled 2025-02-18class ii

GE Medical Systems, LLC: GE Healthcare Revolution CT, System, X-ray, Tomography, Computed

Hazard

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Details from the source

Reason for recall: GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls. Distribution: Worldwide distribution. Quantity: 105 units Lot/code info: System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 480342APEX, UDI (01)00840682146616(11)220900(21)REV2A2200072CN, Mfg. Lot/Serial Number REV2A2200072CN; System ID Number 480342APEX3, UDI (01)00195278460271(11)240600(21)REV2X2400067CN, Mfg. Lot/Serial Number REV2X2400067CN; System ID Number 949250APEX, UDI (01)00195278460271(11)230500(21)REV2X2300008CN, Mfg. Lot/Serial Number REV2X2300008CN; System ID Number 858657APEX, UDI (01)00840682118552(11)160900(21)REVVX1600034CN, Mfg. Lot/Serial Number REVVX1600034CN; System ID Number 310423CT, UDI (01)00840682118552(11)211200(21)REVVX2100047CN, Mfg. Lot/Serial Number REVVX2100047CN; System ID Number 707257APEX1, UDI (01)00195278460271(11)231000(21)REV2X2300077CN, Mfg. Lot/Serial Number REV2X2300077CN; System ID Number 650493REV1, UDI To be provided, Mfg. Lot/Serial Number REVVX1900035CN; System ID Number 408559CTREVO, UDI To be provided, Mfg. Lot/Serial Number REVV81900011CN; System ID Number 904UFNAPEX, UDI (01)00195278460271(11)230300(21)REV2X2300001CN, Mfg. Lot/Serial Number REV2X2300001CN; System ID Number 813972APEX2, UDI (01)00195278460271(11)240700(21)REV2X2400088CN, Mfg. Lot/Serial Number REV2X2400088CN; System ID Number 813972APEX3, UDI (01)00195278460271(11)241000(21)REV2X2400135CN, Mfg. Lot/Serial Number REV2X2400135CN; System ID Number 808522RCT, UDI Not Applicable, Mfg. Lot/Serial Number 00000441716CN8; System ID Number 312996CTRV2, UD

UPC codes

00195278460271, 00840682146616, 00840682118552, 00195278502179, 83195527023, 83155841562, 00840682146319, 00840682123471

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify