← Search recalls
FDA (Device enforcement)Recalled 2025-02-18class ii

GE Medical Systems, LLC: GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed

Hazard

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Details from the source

Reason for recall: GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls. Distribution: Worldwide distribution. Quantity: 4 units Lot/code info: System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 336832APEX, UDI (01)00195278460271(11)241000(21)REV2X2400129CN, Mfg. Lot/Serial Number REV2X2400129CN; System ID Number 973971CTREV, UDI (01)00840682118552(11)160600(21)REVVX1600010CN, Mfg. Lot/Serial Number REVVX1600010CN; System ID Number 513584APEX1, UDI (01)00840682146616(11)230300(21)REV2A2300023CN, Mfg. Lot/Serial Number REV2A2300023CN

UPC codes

00195278460271, 00840682118552, 00840682146616

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify