← Search recalls
FDA (Device enforcement)Recalled 2025-01-27class ii

TAS Medical Inc: TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

Hazard

Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.

Details from the source

Reason for recall: Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation. Distribution: US Nationwide distribution in the states of LA, OK, TN. Quantity: 114 Lot/code info: Model/UDI-DI/Lot(Expiration): T-4000/00197644217267/230403(2024/04/20); T-5000/00860011435921/240801(2025/08/28), 240901(2025/09/25); T-LAP/00860011435914, 00197644927241/240702(2025/07/08), 230906(2024/11/29), 230905(2024/11/29)

UPC codes

00197644217267, 00860011435921, 00860011435914, 00197644927241

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify