The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.
Reason for recall: The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on. Distribution: Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain. Quantity: 3 devices Lot/code info: Software version 2.4.1.0. (1) REF GA-2009996 - Serial number 1791, UDI-DI 07290109145518; (2) REF GA-0006802 - Serial number 1803, UDI-DI 07290109145525; and (3) REF-GA0008700 - Serial number 1818, UDI-DI 07290109145464.
07290109145518, 07290109145525, 07290109145464
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.