← Search recalls
FDA (Device enforcement)Recalled 2025-01-23class ii

LUMENIS, LTD.: (1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser Syst

Hazard

The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

Details from the source

Reason for recall: The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on. Distribution: Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain. Quantity: 3 devices Lot/code info: Software version 2.4.1.0. (1) REF GA-2009996 - Serial number 1791, UDI-DI 07290109145518; (2) REF GA-0006802 - Serial number 1803, UDI-DI 07290109145525; and (3) REF-GA0008700 - Serial number 1818, UDI-DI 07290109145464.

UPC codes

07290109145518, 07290109145525, 07290109145464

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.