← Search recalls
FDA (Device enforcement)Recalled 2023-02-09class ii

DeRoyal Industries Inc: DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

Hazard

The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.

Details from the source

Reason for recall: The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum. Distribution: US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas. Quantity: 27,200 devices Lot/code info: GTIN 50749756377046, Lot 58447585 exp. 12/22/2027

UPC codes

50749756377046

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.